test

This is an old revision of the document!


My Last Words to PXCT Successors:

1) The Turbopump was last used on May 21, 2024. Pumps that have sat for longer than 12 months must be returned to the Agilent factory for rechanneling.

2) The ramp-up speed of the MHV-4 high-voltage supply needs to be set using Putty every time after rebooting the NIM crate. Putty has already been installed on the Laptop desktop. Type command: “SRA 0” for 5 V/s.

3) The half-lives listed in the Safety Office inventory of sources may not always be accurate, so double-check if the source activity is needed in your calculation.

4) All evt, root, and cal data files are in /mnt/daqtesting/pxct
All root and cal files are backed up in /mnt/analysis/e21010/pxct_data

New Purchase / LIBRA2 Upgrade:
1) In 2016, Tamas Budner purchased 27-inch by 12-foot one-sided 1.5-um thick Al-coated foil for GADGET. The remaining foil may be sufficient for use as the PXCT collection foil. Please try to find it. If you can't find it, consider purchasing.

2) In lab 1035, the highest vacuum I've achieved is 1.47E-7 torr. The copper gaskets on the top flange and two side flanges have been reused numerous times. To achieve the high vacuum required by the beamline, it is recommended that new gaskets be purchased.

3) There was a twin of MSD12 manufactured by Micron in the same batch. If we get rich, purchase it as a backup. MSD12 and MSD26 may cost $7k each. Also, consider upgrading to larger area MSD detectors, such as MSD85.

4) The CP5 plus coldhead can be equipped with water-cooled heat sinks instead of the standard internal cooling fans. A closed-loop chiller and external coolant flow lines need to be provided separately.

5) Customer Support Agreement.
https://www.mirion.com/products/technologies/mirion-technologies-services/managed-services-customer-support-agreement/csa-customer-support-agreement
Ship dates: LEGe 06/14/2022; XtRa 07/27/2022;
5 year full warranty on the coldhead;
2 year full warranty on the controller;
2 year full warranty on the detector;

Extended warranty customer support agreement.

PXCT tests July 2023 - May 2024 run_log.docx

XtRa Status Monitor

PXCT Test Reports

pxct_hardware_20230726.pptx
pxct_hardware_20230803.pptx
pxct_hardware_20230808.pptx
pxct_test_20230817.pptx
hardware_setup_for_pxct_20230831.pptx
software_setup_for_pxct_20230831.pptx
pxct_test_20230907.pptx
pxct_20230921.pptx
pxct_traces_noise_diagnostics.pptx
pxct_test_20231005.pptx
pxct_msd_test_20231011.pptx
pxct_test_20231020.pptx
pxct_test_20231026.pptx
pxct_237np_59kev_lifetime_measurement.pptx
pxct_dsl_si_deadlayer_20231109.pptx
pxct_test_20231221.pptx
pxct_test_20240111.pptx
pxct_target_foils_20240118.pptx
pxct_237np_lifetime_test_20240209.pptx
pxct_coincidence_summing_20240221.pptx
pxct_coincidence_test_20240223.pptx
pxct_timing_coincidence_test_20240301.pptx
pxct_daq_deadtime_gamma_eff_test.pptx
pxct_betagamma_20240319.pptx
pxct_60co_test_20240316.pptx
pxct_compton_scattering_test_2020322.pptx
pxct_technical_20240329.pptx
ultra-low-q-value-experiment-feasibility-v0.1.pptx
pxct_low_energy_test_20240405.pptx
ns2024_abstract_dsl_pxct_submitted.pdf
pxct_test_20240412.pptx
pxct_xtra_timing_test_20240417.pptx
pxct_xtra_timing_test_20240418.pptx
bayesian_fit.pptx
pxct_timing_test_20240426.pptx
pxct_xtra_ge_detector_timing_resolution_60co_20240427.pptx
pxct_xtra_ge_detector_timing_resolution_152eu_20240430.pptx
pxct_warm_up_instructions_20240501.pptx
pxct_xtra_ge_detector_timing_resolution_pulser_20240502.pptx
pxct_v10_20240508.pdf
pxct_timing_20240509.pptx
pxct_v11_20240516.pdf
pxct_lege_msd_detector_timing_resolution_241am_20240522.pptx
theoretical_rates_20240609.pptx
lege_x-ray_efficiency_geant4_simulation_20240611.pptx
pxct_v12_20240612.pdf
pxct_v13_20240624.pdf
beta_background_simulation_vs_data_20240628.pptx
59cu_p_α_56ni_reaction_rate_20240702.pptx
lijie_sun_ns2024.pptx
pxct_water_cooling.pptx
lecm2024_lijie_sun.pptx
contingency_plans_for_sauna_scenarios.pptx
contingency_plans_for_warm-up_20240822.pptx
pxct_temp_current_voltage_logs_202306-202408.pptx
pxct_v14_20240825.pdf
pxct_v15_20240905.pdf
pxct_all_geometry_voltages_currents_time_temperatures.pptx
pxct_temp_current_voltage_logs_202306-202409.pptx
all_pxct_applications.pptx
pxct_v16_20240918.pdf
pxct_v19_20241007.pdf
draft_pxct_v21.pptx
interview_nuclear_data_20241030.pptx

refined_pngs_weighting_3class_test_20240627.pptx
refined_pngs_weighting_3class_test_20240702.pptx
64as_pac4_20240805.pptx

doppler_broadening_technique_for_β-delayed_particle_emissions.pptx

get-xia_synchronization_20240530.pptx
get-xia_synchronization_20240531.pptx
e24joe_20240605.pptx
dsl2_eec_s2373_presentation.pdf
dsl2_eec_s2373_proposal.pdf
dsl_s2193_lifetime_sensitivity_estimation_20240514.pptx
weghorn_20240528.pptx
group_meeting_9-13-2024.pptx
e24joe_ddas_run_0150_0303_20240920.pptx
e24joe_228th_runs332_340_room_bkg_run807.pptx
dsl1_emg_fit_2021.pptx
response_function_characterization_for_dsl2.pptx
stopping_power_3he_implantation_layer.pptx
23mg_7787_kev_state_lifetime.pptx
23mg_7787_kev_state_lifetime_bayesian.pptx

Data Acquisition System

LEGe

XtRa

test.1734371554.txt.gz · Last modified: 2024/12/16 12:52 by sunli