design:x-ray_detectors

This is an old revision of the document!


X-rays of interest are 9-70 keV. It is also desirable be able to use an 55Fe source (5.9 keV) for testing purposes. We want to maximize the number of x-rays detected in the previously mentioned range while minimizing background from charged particles and high energy gamma rays. We also have to be able to distinguish between k-alpha x-rays for nuclides with Z and Z+1 where Z >= 28 which corresponds to about 550 eV.

Summary presentation Alex put together for group meeting 7/26/2021 with post meeting updates

Rate calculations spreadsheet

detector_selection_spreadsheet_v1.0.xlsx

kr73_decay_scheme_results_v0.2.pptx

We couldn’t find any detectors which match the energy resolution of LEGE and U-LEGE at low energies. The best we found was ORTEC’s GLP series, shown below. We haven't gotten about the dead layer for the GLP series, so don't know if it's better or worse than LEGe.