GRETINA DAQ system
Status 5/28/2012
Channels:
- Q4P2ε1 - grounded in the detector
- Q6P2ε3 - FET/preamp broken
- Q7P3φ5 - gain 1/10 - broken wire or bad contact in the detector - to be further tested
Radial boxes:
- SN 005207 - over current
- SN 005222 - over current
Spares available: 3
Digitizers:
- SN 067h - errors on bits 9 and 10 with rate >100 kHz
- SN 02Eh - errors on bit 12 with rate ~4 kHz
Spares available: 2 (1 red and 1 blue)
Manual BOOT procedure
- Switch ON the VME crate with the trigger board
- wait 10s
- Switch ON all the VME crates of the digitizers
- wait for green stable lights on the digitizers front panel (~ 2 minutes)
- @op3:
- cd scripts
- ./CheckTS4.sh
- For every digitizer timestamp error reported identify the IOC failing dividing the digitizer number by 4 and adding 1 and (the following example works for IOC#12):
- ssh gsoper:7012@10.0.0.11
- pw:
- push <enter> again
- if you get a prompt like this vme12→
- then give the command <fix
- else push CTRL-x
- wait that the procedure end (~ 2 minutes)
- exit from the IOC control by pushing ~.
- Iterate the previous procedure till ./CheckTS4.sh says that all the timestamp of the digitizers are OK
- cd
- gt
- cd 11-1/gretClust/bin/linux-x86
- $PWD/godaq
- @gretina command line window:
- 2
- 4
- 4b
- 3
- @op3:
- ./setup7modulesRaw.sh
- @gretina command line window:
- 8
- 6
- @gretina DAQ top level
- push cluster
- push nodes
- if not all the processes are green
- then
- delete the red process name (SendRaw)
- push <enter>
- write the process name (SendRaw)
- push <enter>
- push scripts
- push the Set button close to “Setup for Calibration”
- @gretina RUN control
- push start