 CsI Water Damage Repair: CsI Water Damage Repair:
  
-  - Remove CsI crystals **(44/56)**+  - Remove CsI crystals **(56 out of 56)** (11/15)
        * Dismantle telescopes        * Dismantle telescopes
-       * Store E, dE detectors+       * Store E, dE detectors 
        * Unwrap CsI bundles        * Unwrap CsI bundles
-  - Remove as much filter paper as possible from CsI (using ethanol and plastic scraper) **(44/56)** +  - Remove as much filter paper as possible from CsI (using ethanol and plastic scraper) **(56 out of 56)**  (11/20) 
-  - Detach light guide (or photodiode?from each crystal (in refrigerator?) **(1/56)** +  - Document damage (depth on front surface) **(56 out of 56)** 
-  - Design and assemble lapping setup **(work order submitted)** +  - Detach photodiodes from each crystal (leave light guides on for now) **(56 out of 56)** (12/19) 
-       Motion control + support +  - Remove remaining filter paper and light guide paint **(56 out of 56)** (12/19) 
-       Coarse vertical motion +  - MAchine crystals at WashU **(56 out of 56)** (1/18) 
-       CsI holder that attaches to motion control +       * Driven down/back 
-       Sandpaper (ordered+  - Sand sides of each crystal **(56 out of 56)** (3/14) 
-  - Practice lapping with beam blocker +  - Design and assemble lapping setup (4/14) 
 +       Setup machined with some problems 
 +           Lapping machine doesn't fit (FIXED) 
 +           Fast sliders wobbly 
 +       Sandpaper (delivered) (11/18) 
 +       Order stoppers (04/25) 
 +           Spacer (04/28) 
 +  - Clean photodiodes, reattach to light guides 
 +  - Reglue uncoupled light guides 
 +      - more RTV615 needed 
 +  - Practice lapping with worst crystals (4/21) 
 +      - First practice with plastic crystal (04/29) 
 +  - Lap crystals until surface clean (4/28) 
 +  - Polish, rewrap, assemble bundles (5/14) 
 +  - Test individual crystals in dark box (check leakage current) 
 +  - Reassemble telescopes (??)
detectors/csi_repair.1384374671.txt.gz · Last modified: 2013/11/13 15:31 by manfredi