Performing a CSD
- Insert the source FC
- D0738 for ARTEMIS-B; D0717 for HPECR
- Open the Physapps launcher
- Switch to the Apps tab, “All”
- Open Charge State Distribution Scan
- Select source (selection popup may default to the upper left screen)
- Adjust range and gap if desired
- Click
Scan start - Do not close the application until the scan is complete
- When done, click
Save the scanand save in the default folder
Comparing to previous scans
Applying correction
- Use the cursor to get the position of each oxygen peak on the CSD
- Fill in the measured positions in the M/q measured column
- Click
Calcto find the correction - Click
Applyto use it
- Oxygen peaks can be identified as they are the largest ones present
- The measured M/q should be approximately 16/n
SN
- Correction info is vague and needs expanding on, this would heavily improve with a “why” page, but that's a different issue
Showing reference lines