====== Performing a CSD ====== * Insert the source FC * D0738 for ARTEMIS-B; D0717 for HPECR * Open the Physapps launcher * Switch to the Apps tab, "All" * Open Charge State Distribution Scan * Select source (selection popup may default to the upper left screen) * Adjust range and gap if desired * Click ''Scan start'' * Do not close the application until the scan is complete * When done, click ''Save the scan'' and save in the default folder ==== Comparing to previous scans ==== *After scanning, click the ''Open snapshot'' button and select the desired scan *After opening, you can show or hide any loaded snapshots, make sure to select the correct snapshot when using the tools! {{:task:csd_compare.png?}} ==== Applying correction ==== * Use the cursor to get the position of each oxygen peak on the CSD * Fill in the measured positions in the M/q measured column * Click ''Calc'' to find the correction * Click ''Apply'' to use it * Oxygen peaks can be identified as they are the largest ones present * The measured M/q should be approximately 16/n {{:task:csd_correction.png?}} ==== Showing reference lines ====