Line 3: Line 3:
 If MPS tripped due to gate valves closing, please follow the HPECR recovery instructions: If MPS tripped due to gate valves closing, please follow the HPECR recovery instructions:
  
-    *On HPECR operator page, Open Gate Valve FE_ISRC2:BGV_D0675. +   * On HPECR operator page, Open Gate Valve FE_ISRC2:BGV_D0675. 
-    *Go to Microwave Amp. in the 18GHz Microwave Transmitter section.  +   * Go to Microwave Amp. in the 18GHz Microwave Transmitter section.  
-    *On the Microwave Amp page: +   * On the Microwave Amp page: 
-     1. Reset all interlocks / safety interlocks  +       - Check on HPECR Vacuum page and ensure that the BCCG's are okay 
-     2. Turn on PLC commands +       Reset all interlocks / safety interlocks  
-     3. Reset all interlocks / safety interlocks again +       - Turn on PLC commands 
-     4. Turn on BEAM power supply +       - Reset all interlocks / safety interlocks again 
-     5. Reset "RF OUTPUT Set point" to ZERO +       - Turn on BEAM power supply ISC2_HVP_D0652 & ISC2_HVP_D0677 
-     6. Turn ON RF output +       - Turn back on the acceleration column puller PSEL D0678. 
-     7. Recover the RF output power to where it was by 100V/step.  +       Reset "RF OUTPUT Set point" to ZERO 
-    *Go back to the HPECR operator page, turn on both source high voltage and platform high voltage HVP D0652 and HVP D0677. +       - Turn ON RF output 
-    *Turn back on the acceleration column puller PSEL D0678.+       - Recover the RF output power to where it was by 100V/step.  
 +  
 + 