Line 3: | Line 3: | ||
If you are going to change to a low flux around 1E3 pps/cm2, then you'll want to follow the steps below: | If you are going to change to a low flux around 1E3 pps/cm2, then you'll want to follow the steps below: | ||
- | 1. Switch FC_1102 cup to tetram reading on the FC Overview page. | + | - |
- | 2. Calculate the amount of flux needed on FC_2596 by using the calculator on the FSEE GUI. | + | |
- | 3. Measure current on FC_2596 and record the current reading. Do the same on FC_1102. | + | |
- | 4. Calculate the ratio between the FC_1102/ | + | |
- | 5. Put the beam back onto FC_1102 and adjust attenuators/ | + | |
- | 6. Insert the viewer to perform another uniformity measurement. | + | |
- | 7. Send flux to the users /calibrate the jaws. | + | |
- | 8. Beam is ready. | + | |