* [[#General information]] | * [[#General information]] | ||
* [[#Emptying Detectors]] | * [[#Emptying Detectors]] | ||
+ | * [[#Venting the Chamber]] | ||
+ | * [[# | ||
* [[#Filling and Running Detectors]] | * [[#Filling and Running Detectors]] | ||
* [[# | * [[# | ||
Line 22: | Line 24: | ||
===== Emptying Detectors ===== | ===== Emptying Detectors ===== | ||
+ | - Double check that the detectors have been unbiased. | ||
- Set “PPAC Setpoint” to 0 Torr. | - Set “PPAC Setpoint” to 0 Torr. | ||
- Click **EMPTY** on the “GHS_II-PPAC_Main” page to open valves #3 and #4 which will allow for pumping out the gas quickly. | - Click **EMPTY** on the “GHS_II-PPAC_Main” page to open valves #3 and #4 which will allow for pumping out the gas quickly. | ||
Line 37: | Line 40: | ||
To vent the chamber: | To vent the chamber: | ||
- | - Put the GHS in vent mode by clicking **VENT** (valves #5 and #6 are open), | + | - Let the Control Room and the Vacuum Operation Group be aware as EXP alarms may occur, |
+ | - put the GHS in vent mode by clicking **VENT** (valves #5 and #6 are open), | ||
- monitor **RBT_BTS32: | - monitor **RBT_BTS32: | ||
- | - close **RV**, the valve to the roughing pump, and | + | - ensure that **RV**, the valve to the roughing pump, is closed, and |
- open **Vent** valve to start venting. | - open **Vent** valve to start venting. | ||
The PPAC pressure S800_GHS: | The PPAC pressure S800_GHS: | ||
RBT_BTS32: | RBT_BTS32: | ||
- | ===== Evacuating | + | ===== Pumping down the Chamber ===== |
- | {{: | + | Ensure |
+ | * that the chamber is closed | ||
+ | * that the pressure-relief mechanism (on the north side of the chamber, see picture below) is sealing (if not, adjust the flange back in place) | ||
+ | {{ : | ||
+ | * that the manual valve (see picture below) is closed. | ||
+ | {{ : | ||
+ | |||
+ | To evacuate the chamber: | ||
+ | - double-check and make sure that the manual valve is closed, | ||
+ | - put the GHS in pump mode by clicking **PUMP** (valves #3, #5, #6 are open), | ||
+ | - close **FV** below the TMP, and | ||
+ | - open **RV**. | ||
+ | - A message window, which reads, " | ||
+ | - now open the manual valve slowly while watching the local display of the Pirani gauge. Open slowly such that the pressure reading (in Torr) goes down by about 1 unit per second initially. The valve needs to open more as the pressure goes further down. | ||
+ | - On the CSS window, after D2430 turns to yellow indicating that the section is in "POOR VACUUM", | ||
+ | - Turn on TMP, and wait until it reaches the normal rotation (~40,000 rpm). | ||
+ | - Open **TGV**. | ||
+ | - Be sure to go back to S3 and close the manual valve. | ||
===== Filling and Running Detectors ===== | ===== Filling and Running Detectors ===== | ||
- | - If in Vent Mode, click **PUMP** then click **RUN** to enter Run Mode. | + | - If in Vent Mode, click **PUMP** then click **RUN** to enter Run Mode. The valves #1 opens and #3 remains open (#5 and #6 will be closed, and the GHS lines will be isolated from the chamber). |
- If in Pump Mode, click **RUN** to enter Run Mode. | - If in Pump Mode, click **RUN** to enter Run Mode. | ||
- Plot and monitor IG_D2430: | - Plot and monitor IG_D2430: |