| ====== NMR Control ===== | ====== NMR Control ===== | ||
| - | Two digital oscilloscopes are dedicated to NMR readout – one for the analysis line NMRs and one for the spectrograph NMRs. The scopes are located in the S3 mezzanine. They are isolated from clean ground because | + | Two digital oscilloscopes are dedicated to NMR readout – one for the analysis line NMRs and one for the spectrograph NMRs. The scopes are located in the S3 mezzanine. They are isolated from clean ground because |
| ===== Motivation ===== | ===== Motivation ===== | ||
| Line 111: | Line 111: | ||
| * Simply turn it off, wait few seconds, and turn it back on | * Simply turn it off, wait few seconds, and turn it back on | ||
| - | ==== Error message: / | + | ==== Error message: |
| - | * In a new terminal, ssh to yourname@exp-s3-nmr, | + | * In a new terminal, ssh to yourname@exp-s3-nmr, |
| - | * mv spectrograph.log spectrograph.yourname.log | + | * mv spectrograph.log spectrograph-yourname.log |
| * Now close the error message window and you should not have this problem anymore. | * Now close the error message window and you should not have this problem anymore. | ||
| Line 124: | Line 124: | ||
| * This can be alleviated by changing the NMR thresholds (it can be done for each magnet in the analysis line, or for both magnets in the spectrograph). | * This can be alleviated by changing the NMR thresholds (it can be done for each magnet in the analysis line, or for both magnets in the spectrograph). | ||
| * Stop NMR; enter new threshold (in Volts) and click “Set”; start NMR. | * Stop NMR; enter new threshold (in Volts) and click “Set”; start NMR. | ||
| - | * In general, a threshold 0.15 mV is good for normal NMR signals, whereas 0.10 mV may be necessary if the signal | + | * In general, a threshold 0.15 mV is good for normal NMR signals, whereas 0.10 mV may be necessary if the signals |
| 2. Initial field where NMR starts scanning is far away from actual value (this has only been seen with Spectrograph) | 2. Initial field where NMR starts scanning is far away from actual value (this has only been seen with Spectrograph) | ||