This shows you the differences between two versions of the page.
| Both sides previous revision Previous revision Next revision | Previous revision | ||
|
hv_bias [2023/02/13 14:29] noji |
hv_bias [2023/10/17 09:40] (current) noji |
||
|---|---|---|---|
| Line 1: | Line 1: | ||
| ====== Detector bias control ====== | ====== Detector bias control ====== | ||
| - | |||
| ===== HV Remote control ===== | ===== HV Remote control ===== | ||
| - | The high-voltage of the [[Detectors# | ||
| - | The GUI can be opened from the desktop of the u4pc1 computer. Logging in as s800 user, start the singularity container for Debian 8 (Jessie) with | ||
| - | $ jessie_load | + | The high-voltage of the plastic scintillator and the MCP at Object, the [[Detectors# |
| + | |||
| + | {{: | ||
| + | |||
| + | This “S800 HV Bias Power Supply” page is linked from both " | ||
| - | and use the script | + | {{: |
| - | $ / | + | The ramp rate can be changed in the " |
| - | The GUI control panel looks like this when the detectors are biased: | + | Each channel can also be accessed from by going to “FRIB EXP” -> “Power Supplies” and open the “EXP-S3” tab, which looks like the screenshot below. |
| - | {{:wiki:iseg_slot0.png|}} | + | {{:wiki:phoebus_ps_main.png|CS-Studio Phoebus PS HV control}} |
| - | The channel list is found elsewhere. | ||
| The actual HV modules are located on the top level of the S3 vault, under the FP chamber. | The actual HV modules are located on the top level of the S3 vault, under the FP chamber. | ||