This shows you the differences between two versions of the page.
| Both sides previous revision Previous revision Next revision | Previous revision | ||
|
during_experiments [2017/06/17 14:50] pereira [Efficiency of OBJ Scintillator] |
during_experiments [2023/07/20 10:43] (current) swartzj |
||
|---|---|---|---|
| Line 23: | Line 23: | ||
| * Set the trigger to S800 singles and downscale factor 1 | * Set the trigger to S800 singles and downscale factor 1 | ||
| - | * On PanelMate, open page 4 in S800DRIV | + | |
| - | * Click on "upstream mask" | + | |
| + | |||
| + | | ||
| * Start a run without taking to disk and look in the S800 SpecTcl at the 2D xy position spectrum of (s800.fp.crdc1.x, | * Start a run without taking to disk and look in the S800 SpecTcl at the 2D xy position spectrum of (s800.fp.crdc1.x, | ||
| Line 31: | Line 33: | ||
| {{: | {{: | ||
| - | * Remove | + | * Remove |
| - | * Repeat the procedure with CRDC2 by inserting the "downstream mask" | + | * Repeat the procedure with CRDC2 by inserting |
| * {{: | * {{: | ||
| Line 98: | Line 100: | ||
| At relatively high rates (~0.5 MHz), the OBJ scintillator will be gradually damaged. Experimenters should check the efficiency of this detector continuously during the course of the experiment. Whenever the rates drop below ~90%, they may request the Device/Beam Physicist to shim the detector. | At relatively high rates (~0.5 MHz), the OBJ scintillator will be gradually damaged. Experimenters should check the efficiency of this detector continuously during the course of the experiment. Whenever the rates drop below ~90%, they may request the Device/Beam Physicist to shim the detector. | ||
| - | Before shimming, make sure that the thresholds are set properly (not too high). | + | |
| + | |||
| + | Before shimming, make sure that the thresholds are set properly (not too high) (refer to section " | ||