| **Problem 8** | **Problem 8** | ||
| - | The mass resolutions of '' | + | As stated in Problem 7, the mass resolutions of '' |
| - | In order to have a 5% decrease | + | Mres_x = 60.123, and |
| - | (60.123-x)/(60.123) = 0.05 -> x = 57.11685, and (59.262-y)/(59.262) = 0.05 -> y = 56.2989. | + | Mres_y = 59.262. |
| + | |||
| + | We adjusted beam parameters until we saw Mres_x and Mres_y drop to around 55. | ||
| + | |||
| + | The beam position (offsetting | ||
| + | |||
| + | With '' | ||
| + | |||
| + | With '' | ||
| + | |||
| + | The beam size can be changed by adjusting the first and fourth parameter of the '' | ||
| + | |||
| + | With these two parameters each changed from the default 0.0015 to 0.0017, Mres_x | ||
| + | |||
| + | Decreasing these parameters only increases the mass resolution. It seems the smaller the beam spot, the larger the mass resolution. | ||
| + | |||
| + | b) Adjusting the pitch, roll, yaw, x-position, and y-position of Q2 caused a drop in mass resolution at the mass slits (FP2), right after the Wien filter. In COSY, " | ||
| + | |||
| + | {{ : | ||
| + | {{ : | ||
| + | |||